An additional important consideration when measuring laminates, is that
any included air gaps or pockets will cause an early reflection of the
ultrasound beam. This effect will be noticed as a sudden decrease in
thickness in an otherwise regular surface. While this may impede accurate
measurement of total material thickness, it does provide the user with
positive indication of air gaps in the laminate.
Measuring cylinder walls & head ports
Cast iron/aluminum typically vary in consistency from manufacturer to
manufacturer and batch to batch. The nickel content will also vary in cast
iron blocks. Therefore, the user should perform a calibration on the
block/head to be measured. This is commonly done, with a one point
calibration (refer to page 14), by finding an area on the test piece that can
be measured with a set of calipers (Typical locations are at the top of the
block between two cylinders, or any other location with two parallel
surfaces). A two point calibration (refer to page 16) can also be performed
by measuring two points on the test piece, with different thicknesses, that
are representative of the actual range to be tested. For Example, if you
expect to measure a range from .200” - .050”, the hi and low calibration
points should be representative of this range (refer to page 13).